Control Panel
Open care system comes with automatic manual and baby (Servo) modes. In addition the Skin / Air temperature display is large than other set temperature display. Skin temperature display can be converted from ºC to ºF by the push of a button. A big LCD is provided For more information with both audible and visual indicators in the event of :
- High Temp./Low Temp.
- System Fail
- Probe Failure
- Power Failure
- Timer -On
- Heater Fail / ON (Optional)
Heater Assembly
The heater assembly located above the center of the support structure consists of an integral examination light and a 650 watts heating system (Silica or quartz/ceramic heater) for quick warming and even distribution of heat on the baby bed. The heater can be swiveled in either direction for taking X-Ray and doing procedures.
Air Mode (Optional)
The system comes with air sensor. This is added security cover with exclusive feature of switching OFF the heater in case the AIR temperature goes above 39ºC irrespective of what the skin temperature is at the time of event.
Manual Mode
The system automatically starts heating from 40% in manual mode and the same can be changed through settings up to 100%. The system automatically converts from servo mode to manual mode in case of probe failure thus providing nonstop working of the unit.
Baby Bed
The spacious oval baby bed incorporates easy drop down/ fold down glasses for maximum access to the patient. Optional: X-Ray facility is provided below the baby bed for taking X-ray of the baby without disturbing.
The First Breath
The open care system combines virtually all the technology a care giver may need in situation needed for resuscitation. The neonate can be supplied with a constant oxygen flow through the oxygen humidifier when connected to an oxygen cylinder / wall unit. The motorized slow suction can be used with a simple flip of a switch. A digital Apgar time is standard, which signals when 1, 5 and 10 minutes have elapsed.
Castors
The system is fitted with 4″ high quality castors having brakes in front 2 castors. Optional; Antistatic castors
Storage
The unit comes Two Trays below baby bed for keeping consumables.
Phototherapy Unit (Optional)
The Unit can be fitted with LED phototherapy unit or CFL under surface phototherapy unit
Additional Features
If needed the unit can be programmed by user himself for any of the following features:
(A) Count Up / down Timer. (B) Check baby Alarm (C) Pre warming mode